曾华金, 梁会丽, 游静, 李世君, 睢国慧, 屈凌波. 偏最小二乘法结合流动注射化学发光法同时测定芦丁和维生素C[J]. 发光学报, 2013,34(3): 369-374
ZENG Hua-jin, LIANG Hui-li, YOU Jing, LI Shi-jun, SUI Guo-hui, QU Ling-bo. Simultaneous Determination of Ascorbic Acid and Rutin by Flow-injection Chemiluminescence Method Using Partial Least Squares Regression[J]. Chinese Journal of Luminescence, 2013,34(3): 369-374
曾华金, 梁会丽, 游静, 李世君, 睢国慧, 屈凌波. 偏最小二乘法结合流动注射化学发光法同时测定芦丁和维生素C[J]. 发光学报, 2013,34(3): 369-374 DOI: 10.3788/fgxb20133403.0369.
ZENG Hua-jin, LIANG Hui-li, YOU Jing, LI Shi-jun, SUI Guo-hui, QU Ling-bo. Simultaneous Determination of Ascorbic Acid and Rutin by Flow-injection Chemiluminescence Method Using Partial Least Squares Regression[J]. Chinese Journal of Luminescence, 2013,34(3): 369-374 DOI: 10.3788/fgxb20133403.0369.
在碱性条件下
芦丁和维生素C对鲁米诺-铁氰化钾化学发光体系具有显著的抑制作用
但两者峰值出现的时间存在明显差异。基于这一特点
本文将偏最小二乘法与流动注射化学发光法结合
建立了偏最小二乘法-流动注射化学发光同时检测芦丁和维生素C的新方法。在优化实验条件下
发光强度
I
与芦丁浓度和维生素C质量浓度在0.02~0.6 mgL
-1
和0.1~20 mgL
-1
范围内呈线性关系
检出限分别为4.5 gmL
-1
和8.9 ngmL
-1
(信噪比为3)。经实际测定复方芦丁片和人体尿液中芦丁和维生素C的含量表明
该方法具有灵敏度高、分析速度快及操作简便等优点
适合用于药品及生物样品等复杂样品中芦丁和维生素C的测定。
A flow-injection chemiluminescence (CL) system was developed for the simultaneous determination of ascorbic acid (AA) and rutin (RT) using partial least squares (PLS) calibration. This method is based on that fact that both AA and RT can quantitatively reduce Fe
3+
to Fe
2+
and that the reaction rates of AA and RT with Fe
3+
are different. The calibration curve was linear over the concentration range of 0.1~20.0 mgL
-1
and 0.02~0.6 mgL
-1
for AA and RT
respectively. The limits of detection (LOD) for AA and RT were 8.9 gL
-1
and 4.5 gL
-1
respectively
which was calculated according to the International Union of Pure and Applied Chemistry (IUPAC) definition that is 3 times the standard deviation of the blank value. The proposed method was successfully applied to the simultaneous determination of both analytes in pharmaceutical formulations and human urine samples. The results indicate that the method offers the potential advantages of high sensitivity
simplicity and rapidity for AA and RT determination in pharmacy and biological samples.
偏最小二乘法流动注射化学发光芦丁维生素C
flow-injection chemiluminescence (FI-CL)partial least squares (PLS)rutinascorbic acid
National Pharmacopoeia Committee of Ministry of Health. The Drug Standard (Chemical and Pharmaceutical Formulations) [S]. The first volume, Beijing: People's Publishing House, 1989 (in Chinese).[2] Hu K S, Wang S L. Spectrophotometric determination of compound rutinum tablets [J]. Chin. J. Pharm.(中国医药工业杂志), 1990, 21 (2):61-63 (in Chinese).[3] Wang Y, Xiang B R. Simultaneous spectrophotometric determination of rutin and vitamin C in rutin compound tablets with wavelet transform [J]. Phys. Test. Chem. Anal. B (理化检验-化学分册), 2007, 43(10):863-865 (in Chinese).[4] Zhang G F, Miao G. Determination of compound rutoside tablets using high performance liquid chromatography [J]. Chin. J. Chromatogr.(色谱), 1996, 14(3):227-228 (in Chinese).[5] Sun Y H. Determination of vitamin C and rutin in compound rutin tablets by HPLC with dual-wavelength [J]. Anhui Med. Pharm. J. (安徽医药), 2012, 16(1):30-32 (in Chinese).[6] Du D G, Sun S Q, Zhou Q, et al. Quantitative analysis of rutin and vitamin C by NIR FTIR [J]. Spectrosc. Spectr. Anal.(光谱学与光谱分析), 2000, 20(4):474-476 (in Chinese).[7] Chen G, Ding X H, Ye J N. Determination of rutin and L-ascorbic acid in pharmaceutical preparations and fruit juices by capillary zone electrophoresis with electrochemical detection [J]. Chem. J. Chin. Univ.(高等学校化学学报), 2000, 21(9):1348-1352 (in Chinese).[8] Li X J, Xiong H, Yuan Z B. Separation and determination of rutin and vitamin C in rutin compound tablets by capillary electrophoresis [J]. Chin. J. Anal. Lab.(分析实验室), 2001, 20(5):41-42 (in Chinese).[9] Li X J, Zhang Y P, Yuan Z B. Separation and determination of rutin and vitamin C in compound rutin tablets by capillary electrophoresis with amperometric detection [J]. Chin. J. Anal. Chem. (分析化学), 2002, 30(7):815-818 (in Chinese).[10] He F Y, Liu R M, Liu X. Determination of rutin and vitamin C in compound rutin tablets using capillary zone electrophoresis [J]. Chin. J. Anal. Lab.(分析实验室), 2002, 21(4):90-92 (in Chinese).[11] Liu Y D, Ying Y B. The pear acidity quantified analysis using PLS methods and fourier transform near-infrared spectroscopy [J]. Spectrosc. Spectr. Anal.(光谱学与光谱分析), 2006, 26(8):1454-1456 (in Chinese).[12] Cai Z, Zhao J, Jang C Y, et al. Simultaneous determination of 2 constituents in compound rutoside tablets by PLS-UV spectrophotometry [J]. China Pharmacy (中国药房), 2009, 20(31):2454-2456 (in Chinese).[13] Li B X, Wang D M, Xu C L, et al. Flow-injection simultaneous chemiluminescence determination of ascorbic acid and L-cysteine with partial least squares calibration [J]. Microchim. Acta, 2005, 149(3):205-212.[14] Rezaei B, Khayamian T, Mokhtari A. Simultaneous determination of codeine and noscapine by flow-injection chemiluminescence method using N-PLS regression [J]. J. Pharm. Biomed. Anal., 2009, 49(2):234-239.[15] Hasanpour F, Ensafi A A, Khayamian T. Simultaneous chemiluminescence determination of amoxicillin and clavulanic acid using least squares support vector regression [J]. Anal. Chim. Acta, 2010, 670(1):44-50.