The absorption and anisotropy coefficients of the fat emulsion Intralipid-10% were measured. An adding-absorber method was used for measurement of absorption and anisotropy coefficients of the Intralipid-10%. Using the diffusion theory which has been extended to include anisotropy scattering for a point-source in infinite-medium geometry
the standard deviations of multiple measurements for μ
a
/D are less than 0.6% and the residuals between the measured data and the fits have rms values of 0.1%
where μ
a
is the absorbing coefficient of the solution in our experiment and D is the diffusion coefficient of the medium. The values of the standard deviations and the residuals demonstrate the high precision of the measurements