浏览全部资源
扫码关注微信
1. 中国科学院长春光学精密机械与物理研究所应用光学国家重点实验室,吉林 长春,130022
2. 哈尔滨工程大学 理学院,黑龙江 哈尔滨,150001
收稿日期:2013-10-12,
修回日期:2013-12-06,
纸质出版日期:2014-03-03
移动端阅览
彭利荣, 杨兴华, 张健等. 基于光纤的溶解氧浓度荧光测定法[J]. 发光学报, 2014,35(3): 382-386
PENG Li-rong, YANG Xing-hua, ZHANG Jian etc. Fluorescence Detection Method of Dissolved Oxygen Based on Fiber[J]. Chinese Journal of Luminescence, 2014,35(3): 382-386
彭利荣, 杨兴华, 张健等. 基于光纤的溶解氧浓度荧光测定法[J]. 发光学报, 2014,35(3): 382-386 DOI: 10.3788/fgxb20143503.0382.
PENG Li-rong, YANG Xing-hua, ZHANG Jian etc. Fluorescence Detection Method of Dissolved Oxygen Based on Fiber[J]. Chinese Journal of Luminescence, 2014,35(3): 382-386 DOI: 10.3788/fgxb20143503.0382.
针对溶解氧浓度微量探测的现实需求,提出了一种基于荧光猝灭原理、利用多孔光纤实现的溶解氧浓度测定新方法。该方法将钌联吡啶[Ru(dpp)
3
]Cl
2
掺杂的凝胶薄膜修饰在多孔光纤的内壁上,制备了一种溶解氧测定探头并对其测试性能进行表征。光纤贯穿整个长度的孔洞结构既可以作为敏感膜的载体,也可以作为待测物流过的通道和反应场所。与传统测试方法相比,该测试探头的多孔道结构显著提高了比表面积,指示剂可以与溶解氧直接反应,提高了探头的敏感性并且具有微量探测的潜力。实验结果表明,在0~20 mg/L的浓度范围内,Stern-Volmer曲线近似线性,响应敏感度
I
0
/
I
为3.6,响应时间为200 ms。该测试方法在溶解氧微量探测领域具有重要用途。
The paper proposed a new determination method to meet practical needs of trace detection of dissolved oxygen (DO)
which is based on fluorescence quenching and using holy optical fiber (HOF). A DO optical probe was fabricated by modifying the sensitive film doped with fluorophore in the array microholes of HOF
and the sensing character was characterized. The microholes throughout the whole fiber could be used as the substrate of sensing materials and minor reaction pools. Compared with the conventional DO probe
the arrays microholes had huge sensing surfaces and the fluorophore can interact directly with the oxygen molecules
so the sensitivity was improved and also has the potential application in microanalysis detection. From the experimental results
we find the Stern-Volmer plots are linear in the full concentration range of 0~20 mg/L. The ratio of
I
0
/
I
as a sensitivity of the probe is 3.6
and the probe can make a quick response within 200 ms. The probe has important applications in the dissolved oxygen trace detection.
Wang J, Wang L L. Carbon dioxide gas sensor derived from a 547-hole microstructured polymer optical fiber preform[J]. Opt. Lett., 2011, 35(19):3270-3272. [2] Peng L R, Yang X H, Yuan L B, et al. Gaseous ammonia fluorescence probe based on cellulose acetate modified microstructured optical fiber[J]. Opt. Commun., 2011, 284(19):4810 -4814. [3] Egorov A A, Egorov M A, Chekhlova T K, et al. Study of a computer-controlled integrated optical gas-concentration sensor[J]. Quant. Electron., 2010, 38(8):787-790. [4] Chu F H, Yang J J, Cai H W, et al. Characterization of a dissolved oxygen sensor made of plastic optical fiber coated with ruthenium-incorporated sol-gel[J]. Appl. Opt., 2009, 48(2):338-342. [5] Li B, Zuo Q H. Preparation and oxygen-sensing properties of ruthenium (Ⅱ) complex as optical oxygen-sensing material[J]. Chin. J. Lumin.(发光学报), 2011, 32(3):211-215 (in Chinese). [6] Xiong X L, Xiao D, Choi F M M. Dissolved oxygen sensor based on fluorescence quenching of oxygen sensitive ruthenium complex immobilized on silica-Ni-P composite coating[J]. Sens. Actuat. B: Chem., 2006, 117:172-176. [7] Ghasemopur A, Leite A M P, Reynaud F, et al. Hybrid sol-gel planar optics for astronomy[J]. Opt. Exp., 2009, 17(3):1970-1975. [8] Zhao E M, Luo S Z, Yuan L B, et al. Preparation and performance testing optical fiber-optical O2 sensitive probe[J]. Opt. Precision Eng.(光学 精密工程), 2012, 20(11):2411-2415 (in Chinese). [9] Xiong Y, Zhu D, Chen S, et al. A fiber-optic evanescent wave O2 sensor based on Ru(Ⅱ)-doped fluorinated ORMOSILs[J]. J. Fluoresc., 2010, 20(1):269-274. [10] Chu C, Lo Y L. A plastical fiber sensor for the dual sensing of temperature and oxygen[J]. IEEE Photon. Technol. Lett., 2008, 20(1):63-65.
0
浏览量
327
下载量
2
CSCD
关联资源
相关文章
相关作者
相关机构