LIN Jiu-ling, DONG Zhi-shan, ZHANG Xiu-feng, LUO Yong-shi, WANG Zun-li, SUN Tie-zheng, LIN Hai, LIU Xing-ren. A Method for Multi-exponential Fitting to the Fluorescence Decay[J]. Chinese Journal of Luminescence, 2002,23(1): 93-96
LIN Jiu-ling, DONG Zhi-shan, ZHANG Xiu-feng, LUO Yong-shi, WANG Zun-li, SUN Tie-zheng, LIN Hai, LIU Xing-ren. A Method for Multi-exponential Fitting to the Fluorescence Decay[J]. Chinese Journal of Luminescence, 2002,23(1): 93-96DOI:
Among the established techniques for measuring fluorescence decay rate
time-correlated single-photon counting(TCSPC)provides unusually high sensitivity
and is
widely used.Sample's decay curves measured by the TCSPC system contain the system's respond function.In order to get the real decay curve of the sample
we have to do the deconvolution.For that
we need a method to do the nonlinear fitting.By now those established nonlinear fitting method have the problem that resolution is sense to the given initial value
and need too much time to do the calculation.To overcome those disadvantages
we build up a new method to do the fitting.The method is based on the practical pace-accelerated means.It has nice precision and good repetition.For a common value
it can easily work out a fairly good resolution.Using this program
we analyzed some sample's decay curve.The result shows that the new method is useful for the fitting to the fluorescence decay curves.