叶平. 低温光致发光微计算机测试系统[J]. 发光学报, 1986,7(3): 287-294
Ye Ping. PHOTOLUMINESCENCE MEASUREMENT SYSTEM BASED ON AN IBM-PC MICROCOMPUTER[J]. Chinese Journal of Luminescence, 1986,7(3): 287-294
低温光致发光微计算机测试系统(PL-PC)是应用IBM-PC微计算机和先进的具有软件的数据采集系统(Keithley DAS Series 500)等配接到低温光致发光测量设备上
对固体材料的光致发光进行测试.特别是对Ⅲ-V族
Ⅱ-Ⅵ族半导体化合物的测试显著提高了效率和精度.实现了理想的测量控制、数据采集、数据处理和管理.对数据采集采用光编码器等作直接同步控制
采集精度从10Å-0.04Å可调
采集灵敏度<1mV
对测试结果(数据、谱线、计算结果、测试条件等)可以显示、记录、打印、存储等多种方式任选输出
全部过程在程序控制下进行.应用程序使用BASIC和Soft500写成
采用“菜单”方式
模块结构
使用方便
灵活、稳定、可靠.
A low temperature photoluminescence (PL) measurement facility connected with an IBM-PC microcomputer is set up for PL measurement of solid state materials especially for Ⅲ-Ⅴ and
Ⅱ-Ⅴ semiconductor compounds. It is named as PL-PC and as shown in Fig.1. The system is interfaced with a data acquisition system (DAS)
the series 500 of Keithley Company of the United States. Convenient control and efficient data processes are demonstrated in PL-PC experiments.There are many special functions in this PL-PC system for automatic control and convenient use . (1) The spectrum monochromator and chart recorder can be started and stopped automatically by the control of computer key board
or by the designed program. The intensity of input signal can be checked and the output signal would be automatically adjusted by the program. (2) There are different data acquisition abilities
and they can be adjusted to fit varied situations. The resolution of measured spectrum can be chosen as fine as 0.04Å which is good enough for most of grating monochromators in which the resolution is only about 0.1Å. In the case of signal automeasurement
the lowest measurable value can be adjusted to find the fine structure in the PL spectrum. (3) Many data processes are set up by the Series 500. The wavelength and photon energy can be shown and calculated immediately. The intensity of spectrum is corrected according to the spectra responsibility of different detectors
such as the photomultiplier
Ge detector and so on
and can be smoothed to draw out a satisfactory curve. The position of the main peak of luminescence spectra and its full width at the half maximum value are given automatically.(4) The output data can be shown in a variety of ways:data curves on the computer display screen
the printer output and chart recorder curves. Also all of the data can be saved on the floppy disk
which may be used to compare with relevant PL results.In the computer system
soft 500 with basic language is successively used to control and plan all the PL measurement processes
A menu block structure is set up
and all the control orders can be shown on the screen to guide the operator to select and to use this system easily. A PL measurement result of semiconductor HgI
2
with this PL-PC system is given to show the characteristics of output data. This system is helpful for PL experiment and can be expanded to other physical test applications
such as DLTS and optical waveguide measurement.This work was finished in Dept. of ECE
UCSB
U. S. A. Thanks are indebted to Prof. J.L.Merz and the optoelectronics research group for their helps.